Measuring Range Acceleration(ACC) Velocity(VEL)
Displacement(DISP)
H-function
|
(When charge sensitivity 1.0~9.99pc/§¾ is used.) 1, 10, 100, 1000, 10000§¾ , 5-ranges 3Hz~ 1, 10, 100, 1000§¯/s , 4-ranges 10Hz~ 1, 10, 100, 1000§¯/s , 4-ranges 30Hz~ 0.1, 1, 10, 100, 1000§¯/s , 5-ranges 3Hz~ 1, 10, 100, 1000§®p-p, 4-ranges 10Hz~ 0.1, 1, 10, 100, 1000§®p-p, 5-ranges 30Hz~ 0.01, 0.1, 1, 10, 100,§®p-p, 5-ranges Acceleration value by wave detection after passing band pass filter of1kHz. Unit range is the same as acceleration.
|
Filter Characteristic
|
Low pass filter : (LOW.F):OFF(1Hz), 3, 10, 30Hz High pass filter : (HIGH.F):300Hz), 1, 3, 10kHz, OFF,(100kHz) Cut off characteristic : -18dB, butterworth
|
Alarm Function Alarm step Setting range Alarm output
Contact capacitance
|
1-step 0~110% of each range 1a relay contact output individual TTL level output 30VDC, 0.5A, 120VAC, 0.5A
|
Power Supply Battery External battery AC Power supply
|
aD 4-pcs, Life time : 10hrs approx. at continuous use 9 ~ 12V 100V AC ±15%, 50/60Hz, ?20 VA
|